VRN® Dental Ultrasonic Scaler Machine, With LED Detachable Handpiece, Scaling, Perio, Root Canal Irrigation
Beschreibung
Features:
Efficient cleaning: When the ultrasonic scaler machine works, it provides precise voltage frequency to match the vibration frequency of the handle. Under the same power, the handle and working tip will not generate excess heat as energy loss, ensuring the output efficiency, making the scaling process more efficient, faster and comfortable.
High-quality LED handle: The ultrasonic scaler machine handle provides pure white light source illumination with clear vision, can withstand 135℃ sterilization, and titanium alloy material has excellent anti-breakage performance.
High-quality tubing: The internal liquid line tubing and the outer protective layer of the ultrasonic scaler machine are made of high-quality silicone tubing, which is soft and resistant to abrasion.
Parameter:
|
Power input |
100v-240V,50/60Hz |
|
Tip vibration frequency |
28kHz ± 3kHz |
|
Input power |
30VA ~48VA |
|
Ultrasonic output power |
3W~ 20W |
|
Working tips |
G1、G2、G4、P1、E1 |
Packing list:
|
Foot pedal |
1 |
|
PU tube |
1 |
|
Puncture instrument |
1 |
|
Root canal wrench |
1 |
|
Working tips |
5(G1、G2、G4、P1、E1) |
|
HP-4 LED handpiece |
1 |
|
Working tips wrench |
1 |
|
Ultrasonic root canal files |
2 |
|
2-way/3-way connector. |
1 |
|
Power adaptor |
1 |
Trouble shooting:
|
Fault |
Possible cause |
Solutions |
|
The scaling tip doesn’t vibrate and there is no water flowing out when stepping on the foot Switch |
The power plug is in loose contact Make the plug insert to the socket well |
The power plug is in loose contact Make the plug insert to the socket well |
|
The foot pedal is in loose contact(wire foot Switch) |
The foot pedal is in loose contact(wire foot Switch) |
|
|
The fuse in the main unit is broken |
Contact the dealers or us |
|
|
The battery for wireless control foot Switch died |
Change the new batteries |
|
|
The scaling tip does not vibrate but there is water flowing out when stepping on the foot Switch |
The tip is in loose contact |
Screw the tip on the handpiece tightly |
|
The connect plug between the handpiece and the circuit board is in loose contact |
Contact the dealer or us |
|
|
Handpiece problem |
Send the handpiece to dealer or us to repair |
|
|
Cable problem |
Contact the dealer or us |
|
|
The scaling tip vibrates but there is no spray when stepping on the foot Switch |
The water control knob is not on |
Turn on the water control knob |
|
There is impurity in the electric-magnetic valve |
Take the electric-magnetic valve apart or contact the dealer or us |
|
|
The water system is blocked |
Clean the water pipe by multi-function syringe |
|
|
There is still water flowing out after the power is off |
The electric-magnetic valve problem |
Contact the dealer or us |
|
The handpiece generates heat |
The water control knob is in a low setting |
Turn the water control knob to a higher qrade |
|
The amount of spouting water is too little |
The water control knob is in a low setting |
Turn the water control knob to a higher grade |
|
The water pressure is not high enough |
Make the water pressure higher |
|
|
The water system is blocked |
Clean the water pipe by multi-function syringe |
|
|
The vibration of the tip becomes weak |
The tip has not been screwed on to the handpiece tightly |
Screw the tip tightly |
|
The coupling between the handpiece and the cable is not dry |
Dry it by the hot air |
|
|
The tip is damaged |
Change the new scaling tip |
|
|
There is water leaking from the coupling between the handpiece and the cable |
The waterproof O-ring is damaged |
Change the new waterproof O-ring |
|
The U-file does not vibrate |
The screw is loose
|
Tighten it |
|
Endo chuck is damaged |
Change the new endo chuck |
|
|
There is noise coming from the endo chuck |
The screw is loose |
Tighten it |
|
The LED doesn't work |
LED in loose contact |
Check the circuit |
|
LED has burned out |
Change the LED |